Chemical Purity D-Luciferin
QUALITY CONTROL PROCEDURES - HPLC analysis of chemical purity
The biochemical performance of each lot of D-luciferin is assayed at 25°C by an automatic procedure using
a special quality control kit containing standardised preparations of all components (luciferase, ATP, buffer etc.)
required for the measurement.
Light emission is measured at the following final concentrations of D-luciferin: 0.1 mg/mL (below optimum),
0.2 mg/mL (optimum) and 0.3 mg/mL (above optimum). Each run begins and ends with 0.2 mg/mL of a reference lot of D-luciferin.
Assays are performed in triplicates for each sample.
The analysis is performed with Shimadzu SPD-10A UV Spectrometric Detector in Shimadzu Liquid Chromatograph. The procedure uses a buffered-reverse phase technique with a gradient elution. An example of a chromatogram obtained with Lot 1035 at 330 nm is shown in Fig. 1. Results from 5 samples performed in duplicates are summarized in Table 1. Results obtained at 265 nm were similar. In the interval 220-540 nm no additional chromatographic peaks were detected. The average retention time of D-Luciferin was 7.126±0.013 min.
Peak no.2 at 9.302±0.021 min corresponds to deoxyluciferin. The average percent purity for Lot 1035 is 99.65±0.03% D-Luciferin. According to specifications, the percent purity of D-Luciferin (free acid) must be greater than 99.5%.
HPLC analysis of chemical purity
Fig. 1: Chromatogram of D-Luciferin (Lot 1035) at 330 nm. Insert shows absorbance using an enlarged scale.

Table 1: Chemical purity analysed by HPLC. UV detection at 330 nm
| Retention time | Area (%) |
| 7.126 | 99.68 |
| 9.302 | 0.02 |
| 16.748 | 0.10 |
| 19.321 | 0.06 |
| 20.599 | 0.14 |
